荧光光谱分析仪的使用方法:
The method of using the fluorescence spectrum analyzer:
开机后运行 tcm4400 按以下顺序进行: 20kv/10ma→30kv/10ma→40kv/10ma→50kv/10ma→60kv/10ma→24kv/100ma→40kv/60ma→50kv/48ma →60kv/40ma 如仃机时间大于 24 小时小于 100 小时,每步停留时间为 1 分钟。如仃机时间大于 100 小时,每步停留时间为 5 分钟。 自动老化开机后运行 tcm4400,如仃机时间大于 24 小时小于 100 小时,选择“fast”老化,如仃机时间大于 100 小时,选择“normal”老化。
After the start of operation of tcm4400 in the following order: 20kv/10ma - 30kv/10ma - 40kv/10ma - 50kv/10ma - 60kv/10ma - 24kv/100ma - 40kv/60ma - 50kv/48ma - 60kv/40ma for machine time more than 24 hours less than 100 hours, each residence time is 1 minutes. As for time more than 100 hours, each residence time is 5 minutes. Automatic aging running tcm4400 boot, such as stopping the machine for more than 24 hours less than 100 hours, the selection of "fast" aging, such as stopping the machine for 100 hours, the selection of "normal" aging.